FUNCTIONS OF KEYS, CONTROLS, INDICATIONS, AND CONNECTORS

1

On / Off / Tune Switch

ON position - energizes the power supply.
OFF position - de-energizes the power supply.
TUNE position - energizes the power supply
momentarily for tuning purposes.

2

Pilot Light

Illuminates when the power supply is energized.

3

Continuous / Pulsed *

Sets the duty cycle to continuous or pulsed mode.

4

% Duty Cycle Selector *

Sets the pulse rate when CONTINUOUS / PULSED
Switch is set to pulsed. Intermittent operation inhibits
heat build-up in the solution and provides more efficient
processing by allowing the material to settle back under
the probe tip after each burst.

5

Timer

Sets the duration of ultrasonic application (up to 15
minutes). In the HOLD position, the timing mode is
uninhibited and the ultrasonic energy is applied
indefinitely.

6

Output Control

Controls the amplitude of vibration at the probe tip.

CAUTION

When using a microtip, never allow the Output control
to exceed MICROTIP LIMIT “5”.

7

Power Monitor

Indicates the percentage of ultrasonic power delivered to
the probe.

8

3 Pin Connector

Connects to the converter cable.

9

3 Pin Connector **

Connects to the converter cable (available only with 500
watt dual output units).

10

Tuning Control

Optimizes performance by matching the frequency of
the power supply to that of the converter / probe
assembly

11

Fuse

Protects against electrical overload.

12

Electrical Line Cord

Connects the power supply to t he electrical outlet.

* Available only with Ultrasonic Processors equipped with pulsers.