AMPLITUDE:

The amplitude is the only parameter that must be set in order for

the Ultrasonic Processor to be operational. The other control parameters – Time
and Pulse, do not have to be set for continuous operation. AMPL. displays the
percentage of maximum of amplitude e.g. 40%, set by the AMPLITUDE control.
Rotate the AMPLITUDE control for a 40% reading on the LCD screen – Ampl
40%.

The screen will display:






The Ultrasonic Processor is now ready for continuous operation. To energize ultrasonics,
press the

START

key or the footswitch. To de-energize ultrasonics, press the

STOP

key

or release the footswitch. If the Time, Temperature, Pulse, Save, or Recall functions must
be used, refer to the appropriate paragraph(s) below.


TIME __:__:__ TEMP __ __

°C

PULSE__:__:__ AMPL 40 %

NOTE

Any combination of functions can be selected in any order. To clear an erroneous
entry press the

CLEAR

key.

NOTE

If the

START

key is pressed and the time limit has not been set, processing will remain

uninterrupted until the

STOP

key is depressed.


If the

START

key is pressed and the time limit has been set, processing will remain uninterrupted

until the set time limit expires, or the

STOP

key is pressed – whichever occurs first.


If a footswitch is use, and the time limit has not been set, processing will remain uninterrupted as
long as the footswitch is depressed.

If a footswitch is used, and the time limit has been set processing will remain uninterrupted until
the time limit expires or the footswitch is released – whichever occurs first.

The

START

key and footswitch are mutually exclusive. If the process is initiated by the

START

key, the footswitch becomes inoperative. If the process is initiated by the footswitch, the

STOP

key becomes inoperative.