23


5.

Immerse the temperature Probe in the center of the ice water bath for a period
of 40 seconds. Do not allow the probe to contact the vessel. When the self-
calibration for low temperature is complete, the screen will display the
following message:


6.

Immerse the Temperature Probe in the center of the boiling water for a period
of 40 second. Do not allow the probe to contact the vessel.

When the calibration for high temperature is complete, the screen will display the
following message:



















TEMPERATURE PROBE CALIBRATION

PLACE TEMPERATURE PROBE

INTO ICE WATER BATH

PLACE TEMPERATURE PROBE

INTO BOILING WATER

TEMPERATURE PROBE

CALIBRATION COMPLETED