FUNCTIONS OF CONTROLS, INDICATIONS AND CONNECTORS

POWER switch


When depressed, applies electrical power to the unit

Illuminates in the ON position.

TEST switch

Used to determine if the system is working properly. With the
AMPITUDE control set at 70, and the probe in air (not in the
liquid), proper operation will be demonstrated by a reading of
20% or less on the power monitor when the

TEST

switch is

depressed. Make sure that the liquid has been drained out of
it. If reading exceeds 20%, check the probe for excessive
erosion, and the converter for elevated temperature.

RESET switch


If an overload condition should occur the

RESET

switch will

illuminate. Depress the

RESET

switch to reset the power

supply.

TIMER


Sets the duration of ultrasonic applications.

AMPLITUDE

control


Controls the amplitude of vibrations at the probe tip.

Power monitor


Indicates the percentage of ultrasonic power delivered top the
probe. (e.g. 50% = 750 watts)

Converter cable


Connects the power supply to the converter.

Fuse


Protects against electrical overload.

Power cord


Connects the power supply to the electrical outlet.