Preparation for use, Preparations for use – Sonics VCF1500HV User Manual

Page 8

Advertising
background image



PREPARATION FOR USE

NOTE

The High Volume Flow Cell is recommended only for the treatment of low viscosity
samples, which do not require extended exposure to ultrasonics. Because the residence
time within the cell is relatively short, it might be necessary to recirculate the sample.

CAUTION

Do not operate an Ultrasonic Processor that has been transferred
from a very cold environment to a hot environment, or vice versa.
Wait until it has reached room temperature

1.

Ensure that the AMPLITUDE dial is set fully counter-clockwise.


2.

Using the spanner wrenches provided, secure the converter to the High Volume Flow
Cell.


3.

Connect the converter cable to the power supply.

CAUTION

Never place a washer between the probe and the converter.
Never apply grease to the mating surfaces or threads of the
converter or probe.

4.

Ensure that the

POWER

switch is set to OFF.


5.

Plug the electrical line cord into the electrical outlet.

6.

Drain the High Volume Flow Cell completely.


7.

Set the AMPLITUDE to 50.

8

Go To Top Of Document

Advertising