Operating conditions, Safety during operation, Notice – VACUUBRAND DVR 3 User Manual

Page 5

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual: 999072 / 12/01/2011

page 5 of 18

• Comply with

max. permitted ambient and gas temperature (see ”Technical

data”) and make sure ventilation is adequate if the equipment is installed in a

housing or if ambient temperature is elevated.

• Avoid high heat supply (e. g. due to hot process gases).

• In case of residues or aggressive or condensable media install a gas washing

bottle if necessary.

When the device is brought from cold environment into a warm room for operation

bedewing may occur. Allow the device to acclimatize.

Comply with

national safety regulations and safety requirements concerning the

use of vacuum and electrical equipment. Comply with all

applicable and relevant

safety requirements (regulations and guidelines), implement the required ac-

tions and adopt suitable safety measures.

Operating conditions

DVR 3 has only an approval according to device labelling.

Ensure that the materials of the wetted parts are compatible with the substances

processed in the vacuum system, see section “Technical data”.

Safety during operation

Adopt suitable measures to prevent the release of dangerous, toxic, explosive,

corrosive, noxious or polluting fluids, vapors and gases.

Prevent any part of the human body from coming in contact with the vacuum.

Max. permitted pressure at the vacuum gauge: 1.1 bar absolute with hose nozzle;

1.5 bar absolute with small flange connection.

The display flashes at a pressure above 1060 mbar.

Attention: If the pressure is higher than approximately 1060 mbar the pressure

reading becomes incorrect (saturation of the pressure transducer). The display

flashes.

Immediate pressure relief necessary! Risk of bursting!

• Use only

genuine spare parts and accessories. Otherwise safety and perfor-

mance of the equipment as well as the electromagnetic compatibility of the equip-

ment might be reduced.

Possibly the CE mark becomes void if not using OEM spare parts.

Electronic equipment is never 100% fail-safe. This may lead to an ill-defined sta-

tus of the equipment or of other connected devices. Provide protective measures

against malfunction and failure. Ensure that in case of failure the device and the

vacuum system always will turn into a safe status.

NOTICE

NOTICE

NOTICE