Electrical Connections

CAUTION: Do not use a wrench to tighten the SMC con-
nectors on the bias and electrometer cables.
Connections should be finger tight only.

1. Referring to

Figure 4 as necessary, connect BIAS-B cable to the top

electrode and the electrometer cable (

ELECT) to the bottom electrode.

The

BIAS-A remains disconnected.

2. Connect the high-voltage cable from the detector to the pulse supply, and

connect the pulse supply cable between the back of the controller and
the pulse supply. Connect the heater cable from the detector to the back
of the controller.

CONTROLLER

(Rear panel)

HEATER CABLE

FROM DETECTOR

PULSER

MODULE

HIGH

VOLTAGE

ELECTROMETER

BIAS-A

BIAS-B

RED
BLACK
WHITE
GREEN
BROWN
BLUE
ORANGE
YELLOW

RED
BLACK
GREEN
WHITE

0 - 10 V
Common
0 - 1 V
Shield
COM
SET

UNATTENUATED

OUTPUT

AUTO

ZERO

TO COMPUTER

OR INTEGRATOR

+

ATTENUATED

OUTPUT

SHIELD

TO

RECORDER

(no connection)

Figure 4: Electrical connections