Initial power-up, Troubleshooting high background current – VICI D-4-I User Manual

Page 21

Advertising
background image

17

Initial Power-Up

1. Set the

MODE switch on the front of the controller to PDHID.

2. Set the discharge gas flow as specified in on page 13 in the section

entitled “Connecting the Discharge Gas to the Detector”.

CAUTION: Always make sure that discharge gas is
flowing before powering up the detector.

3. Apply power to the helium purifier.

4. Turn on the

MAINS switch located on the back of the controller and

the

DISCHARGE switch on the front of the controller. The discharge

should start within five minutes. (Once a system has been up and
unning, the discharge will start within a few seconds.) In a clean
system the discharge will have a peach/pink color. A purple discharge
is an indication of impurities and/or leaks in the discharge gas stream.

5. Set the detector temperature with the

TEMPERATURE control knob.

6. Check the detector standing/background current, which is indicated in

the LED

DISPLAY on the controller. The optimum detector background

current is in the range of 1.0 to 2.5 nA. The initial value may be higher,
but as the detector bakes out at its operating temperature, the back-
ground current should decrease to the optimum value.

If the standing current reaches an acceptable level, the detector is ready

for use. Proceed to page 18 – “Mode Selection and Setup”.

Troubleshooting High Background Current

If the background current does not drop below 2.5 nA even after a 12 hour
bakeout, there is either a leak in the system or the column effluent is not
clean. To see if the high background current is due to the column:

1. Make sure the controller is in the un-zeroed condition. (Refer to the

discussion about the ZERO push button on page 7.)

2. With a capillary column, loosen the knurled nut and pull the column out

~20 mm. Secure the nut.

With a packed column, completely disconnect the column from the

column inlet tube, leaving the inlet open.

3. Watch the detector standing/background current, indicated in the control-

ler

DISPLAY. If the current remains high, then either the system has a

leak in the discharge gas supply line or the discharge gas has impurities
in it. Proceed to the next section, “Checking for Leaks in the Discharge
Gas Plumbing”. If the current decreases dramatically, then either the
carrier gas supply has leaks and/or contaminants, or the column is the
source of contamination and needs a bakeout. Read the “Column
Bakeout Precautions” on the next page before proceeding.

Installation

Advertising