Installation

Pulser Module Installation

1. Mount the pulser module in the pneumatics area of the GC, using existing

slotted screw holes. (Figure 5)

2. Power to the pulser module is provided through cable assembly I-23628.

Drill a 12.0 mm diameter hole in the front panel of the GC to mount the
on/off switch.

3. Unscrew the switch mounting nut and remove it from the cable assembly

by sliding it over all the wiring.

4. Feed the wiring through the 12 mm hole from the front of the GC until the

threaded portion of the switch follows the wiring through the mounting hole.

5. Slide the switch mounting nut back over all the wiring, and tighten the nut

onto the threads.

6. Install the 4-pin connector by

inserting Pins 1-4 of the cable
assembly into the connector as
shown in Figures 6 and 7. (Once a
pin is snapped into the connector,
it cannot be removed.)

7. Plug the 4-pin connector into the pulser module.

8. Figure 8 is a schematic of a portion of the Varian GC power board.

Connect the red wire (+) of the I-23628 cable to pin 14 of “External
Events” connector J43, and the black wire (-) to pin 13.

1 2 3 4 5 6 7 8

9 10 11 12 13 14

TRANSFORMER

JACK

CHASSIS FAN
JACK

J42

J43

EXTERNAL

EVENTS

9 10 11 12 13 14

BLACK
RED

MAIN POWER

HARNESS JACK

COLUMN OVEN

HEATER JACK

Figure 8: Connections for 24 VDC for the pulser module

BLACK

BLACK

GREEN

RED

BLACK

BLACK

+

-

RED

BLACK

PIN 1

PIN 2

PIN 4

PIN 3

SWITCH MOUNTING NUT

Figure 6: Pulser cable assembly

BLACK 4
BLACK 2

3

GREEN

1

BLACK

Figure 7: 4-pin connector