5.2.4.

Minimum Division (Weighing Interval)

Step 4

Step 5

Check the |

d-xx

I

display. The xx is minimum division (01,02, 05,10, 20,50).

Minimum division

Select a minimum division using the following keys,

key... The key to select a parameter,

key... The key to store a parameter and

proceed to step 6.

A

mode

! key..The key not to change a parameter

and proceed to step 6.

M-H key ..... The key to proceed to "5.6. Gravity

Compensation Function".

To Gravity

Compesation Function

To Maximum Capacity

5.2.5.

Maximum Capacity

>

Step 6 Check the [ ixxxx | display. The xxxx is current

Maximum capacity

capacity. Select a new maximum capacity

using the foliowing keys.

key......... . The key to select a figure.

key...........The key to select a number.

key.......... The key to store a parameter

and proceed to step 7 (of the

zero calibration).

A

I

mod

^ key.......... The key not to change a

To Weighing Range

parameter and proceed to step 7

Punction

(of the zero calibration).

To Zero Calibration

PHESETTABg key,,..... The key to porceed to "5.3.

Weighing Range Function".

5. Calibration

Page 14

AD-4329 Weighing Indicator