3. Installation and Precautions

ЗЛ .1. Instailation and Precautions

Aa

A

d

The AD-4329 weighing indicator is a precision electronic instrument Handle the

Indicator carefully.

The operating temperature is -10t: to *f40t: (14“F to 104"^).

Do not install the scale in direct sunlight.

Misoperation or other problems may be caused by an unstable power source including

momentary power failrue or instantaneous noise. Use a stable power source.

Do not connect the power cord before the installation Is finished.

Please confirm that the local voltage and receptacle type are correct for your scale.

The jOperate/Standbvi key @ key) can not cut the power. The load cell and the internal

circuit are supplied with power in the standby state.

Use shielded cable for ail connections. Connect the cable shields to the shield

terminal or case as an earth terminal.

Earth ground the indicator. Do not Join the earth ground line with other electric power

equipment. There is an earth ground terminal at the power cord receptacle.

Do not install the scale in a place where it is af3t to be charged with static electricity,

or where the relative humidity is lower than 45%RH. Plastic and isolators are apt to be

charged with static electricity.

3.1.2.

The Load Cell Connections

Connect the load ceil wires to the connector (receptacle), at the rear panel, using the

accessory load cell plug.

It is possible to connect a 4 wire cable that 1 pin"2pin and 3pin-4pin are shorted, if the

distance between the indicator and a load cell is shorter than 5m.

The output voltage of a load ceil is a very sensitive signal. Space the load cell cable

away from any noise source.

It is possible to connect eight 350Q load cells.

The load cel! drive is 5VDC ± 5% between EXC+ and EXC-, the maximum current 120mA.

Cable

Cable

EXC+

EXC-

SIG-

S1G+

Load ceii

i\

Positive Sense

Positive Excitation Input

Negative Excitation input

Negative Sense

Negative signal Output

Positive signai Outpur

Load Cell Plug

Standard
connection

AD-4329 Weighing Indicator

Page 5

Available connection for

a cable shorter than 5m.