Hanna Instruments HI 3512 User Manual
Page 62

62
CHRxx
Change the instrument range according with the parameter
value (xx):
• xx=10 pH range/0.001 resolution on channel 1
• xx=11 pH range/0.01 resolution on channel 1
• xx=12 pH range/0.1 resolution on channel 1
• xx=13 mV range on channel 1
• xx=14 Relative mV range on channel 1
• xx=15 ISE range on channel 1
• xx=20 EC range on channel 2
• xx=21 Resistivity range on channel 2
• xx=22 TDS range on channel 2
• xx=23 Salinity range on channel 2
The instrument will answer for these commands with:
<STX> <answer
> <ETX>
where:
<STX> is 02 ASCII code character (start of text)
<ETX> is 03 ASCII code character (end of text)
<answer>:
<ACK> is 06 ASCII code character (recognized command)
<NAK> is 21 ASCII code character (unrecognized com-
mand)
<CAN> is 24 ASCII code character (corrupted command)
COMMANDS REQUIRING AN ANSWER
The instrument will answer for these commands with:
<STX> <answer> <checksum> <ETX>
where the checksum is the bytes sum of the answer string sent as 2 ASCII
characters.
All the answer messages are with ASCII characters.
RAS
Causes the instrument to send a complete set of readings in
according with the current range:
• pH, mV reading and temperature on pH range.
• mV and temperature reading on mV range
• Rel mV, absolute mV and temperature reading on
Rel mV range.
• concentration, mV and temperature reading on ppm
range.
• EC and temperature reading on the EC range