TAG INSTALLATION

The tag is housed in a rugged metal that can withstand harsh environments. However, it is better
to protect the tag from direct rain.
Place the tag near a sampling point. Fix it securely with the provided screws, in such a way that
the metallic iButton

®

is easily accessible for reading the tag.

The number of tags that can be installed is practically unlimited. Additional tags can be ordered
(

HI 920005 - five tag holders with tags).

LAMP REPLACEMENT

The instrument tungsten lamp has a life longer than 100,000 measurements. In case of lamp
failure, the defective lamp can be easily replaced. When the lamp is broken, the instrument displays
“

no L” error message.

To replace the lamp follow the next steps:
• Remove the battery lid.
• Unscrew the lamp connection using a

screwdriver.

• Unlock the lamp and extract it by pulling it out

from the lamp holder handler.

• Place the new lamp in the right position and

push it until is securely locked.

• Insert the lamp leads into the connector and

tight them using a screwdriver.

Warning: After lamp replacement the meter has

to be recalibrated.