Hanna Instruments HI 88713 User Manual

Page 12

Advertising
background image

22

23

Note: The reading of the first point could be skipped for FNU and NTU modes by pressing “Skip”
functional key. In this case, the 0.00 value will be used for calibration.

•

Place the cuvette for first calibration point:

•

<0.1 FNU (NTU) calibration cuvette for all

modes except FAU mode.

•

15 NTU formazin prepared standard for FAU

mode.

Note: For the FAU mode, if necessary, press UP
or DOWN key to edit the calibration point value
to match the exact value of the standard.

•

Close the lid and press “Read” functional key. The display will show the value blinking and the

light icon during measurement.
Note: If the Average mode was previously selected, the measurement in calibration mode will be
done using the average.

•

At the end of the measurement, the second proposed calibration point is displayed.

SECOND POINT CALIBRATION
•

Remove the first calibration cuvette.

•

Place the cuvette for second calibration point:

•

15 NTU (FNU) calibration cuvette for NTU

ratiometric mode.

•

15 NTU formazin prepared standard for NTU

non-rating and FNU modes.

•

100 NTU formazin prepared standard for FAU mode.

Note:

If necessary, press UP or DOWN key to edit

each calibration point value to match the exact value of

the standard as measured with a reference turbidimeter.

•

Close the lid and press “Read” functional key.

The display will show the value blinking and the light icon during

measurement.

•

At the end of the measurement, the third proposed calibration point is displayed.

•

At this moment it is possible to exit calibration by pressing “End” functional key.

•

If the calibration is terminated, the display will briefly

show “Store...” and the two point calibration is saved.
The instrument returns to the main screen.

THIRD CALIBRATION POINT
•

Remove the second standard cuvette.

•

Place the cuvette for third calibration point:

•

100 FNU (NTU) calibration cuvette for NTU

ratiometric mode.

•

100 NTU formazin prepared standard for NTU

non-ratio and FNU modes.

•

750 NTU formazin prepared standard for FAU mode.

Note:

If necessary, press UP or DOWN key to edit each

calibration point value to match the exact value of the
standard as measured with a reference turbidimeter.

•

Close the lid and press “Read” functional key.

The display will show the value blinking and the light icon during

measurement.

•

At the end of the measurement, the fourth proposed calibration point is displayed.

•

At this moment it is possible to exit calibration by pressing “End” functional key.

•

If the calibration is terminated, the display will briefly

show “Store...” and the three point calibration is saved.
The instrument returns in the main screen.