Hanna Instruments HI 88713 User Manual
Page 5

8
9
Note: If you are using multiple cuvettes, always match the cuvettes.
CUVETTE HANDLING
The cuvettes should be free of scratches or cracks. Any cuvette with visible scratches should be
discarded. The cuvettes should be periodically washed with acid. After washing, the cuvettes
should be rinsed well with distilled or deionized water, multiple times. Allow cuvettes to air-dry
and store them for long periods of time with theire caps, to avoid dirt entering inside. Always
handle the cuvette by touching only the cap or its top side (over the horizontal line).
Always store the cuvettes in separate boxes or with separators between them to avoid scratches
on the surface.
CUVETTE PREPARATION
Whenever a cuvette is used, it must be clean inside and outside. When it is placed into the
instrument, it must be dry outside, completely free of fingerprints or dirt.
GENERAL TIPS FOR AN ACCURATE MEASUREMENT
GENERAL TIPS FOR AN ACCURATE MEASUREMENT
GENERAL TIPS FOR AN ACCURATE MEASUREMENT
GENERAL TIPS FOR AN ACCURATE MEASUREMENT
GENERAL TIPS FOR AN ACCURATE MEASUREMENT
The HI 88713 is a highly accurate bench too turbidity meter. Special care must be taken during
calibration and sample preparation to optimize the instrument performance and to fully benefit from
its features. Is very important for the analyst to use proper measurement techniques for accurate,
precise and repeatable readings. The instructions listed below should be carefully followed during
calibration and measurement to ensure optimum accuracy.
GENERAL RULES
• Always put the instrument on a flat, rugged surface when taking measurements.
• Do not operate in direct sunlight.
• Keep the lid of the instrument closed when it is not used to prevent dust or dirt entering inside.
• Always close the cuvette lid during measurement.
• Always use cuvettes without scratches or cracks because they can cause inaccurate readings.
• Always cap the cuvettes to prevent the sample from spilling
• To prevent contamination of the optical system, do not use to much oil.
• When using multiple cuvettes, they should be indexed and matched, when possible.
CUVETTE
The cuvette is part of the optical system in all measurements. The light reaches the sample by
passing through the cuvette glass. As a result, the measurement can be affected by the glass
imperfections, dirt, dust, scratches, or fingerprints present on the cuvette surface. Special care
must be taken in preparing and handling the cuvette.
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
FNU mode
Range
0.00 to 9.99; 10.0 to 99.9; 100 to 1000 FNU
Resolution
0.01; 0.1; 1 FNU
Accuracy
±2% of reading plus straylight
FAU mode
Range
10.0 to 99.9; 100 to 4000 FAU
Resolution
0.1; 1 FAU
Accuracy
±10% of reading
NTU ratio mode
Range
0.00 to 9.99; 10.0 to 99.9; 100 to 4000 NTU
0.00 to 9.99; 10.0 to 99.9; 100 to 980 EBC
Resolution
0.01; 0.1; 1 NTU
0.01; 0.1; 1 EBC
Accuracy
±2% of reading plus straylight
±5% of reading above 1000 NTU
NTU non ratio mode
Range
0.00 to 9.99; 10.0 to 99.9; 100 to 1000 NTU
0.00 to 9.99; 10.0 to 99.9; 100 to 245 EBC
Resolution
0.01; 0.1; 1 NTU
0.01; 0.1; 1 EBC
Accuracy
±2% of reading plus straylight
Range selection
Automatic
Repeatibility
±1% of reading or straylight, whichever is greater
Stray Light
< 0.1 NTU (0.05 EBC)
Light Detector
Silicon Photocell
Method
ISO 7027 Method
Measuring mode
Normal, Average, Continuous.
Turbidity Standards
<0.1, 15, 100, 750 FNU and 2000 NTU
Calibration
Two, three, four or five-point calibration
Light Source
IR LED
Lamp life
Instrument life
Display
40 x 70mm graphic LCD (64x128 pixels) with backlight
LOG Memory
200 records
Serial Interface
USB
Environment
0°C (32°F) to 50°C (122°F); max 95% RH non-condensing
Power supply
12 Vdc power input
Dimensions
230 x 200 x 145 mm (9 x 7.9 x 5.7”) L x W x H
Weight
2.5 Kg (88 oz.)