GENERAL DESCRIPTION

ABBREVIATIONS

SPECIFICATIONS

PRECISION AND ACCURACY

The HI 96717 is an auto diagnostic portable microprocessor meter that benefits from Hanna’s
years of experience as a manufacturer of analytical instruments. It has the advanced optical
system based on a special tungsten lamp and a narrow band interference filter that allows most
accurate and repeatable readings. All instruments are factory calibrated and the electronic and
optical design minimizes the need of frequent calibration.
With the powerful CAL CHECK

TM

validation function, you are able to validate good performance

of your instrument at any time. The validation procedure is extremely user friendly. Just use the
exclusive HANNA ready-made, NIST traceable standards to verify the performance of the
instrument and recalibrate if necessary.
All instruments are splash waterproof and the lamp and filter units are protected from dust or dirt
by a transparent cup. This makes the instruments fulfill field applications. Display messages aid
the user in routine operation. The meter has an auto-shut off feature that will turn off the
instrument after 10 minutes of non use in

measurement mode or after 1 hour if left in

calibration mode.
The meter uses an exclusive positive-locking system to ensure that the cuvette is in the same
position every time it is placed into the measurement cell. It is designed to fit a cuvette with a
larger neck making it easier to add both sample and reagents. The cuvette is made from special
optical glass to obtain best results.
The HI 96717 meter measures the phosphate (PO

4

3-

) content in water samples in the 0.0 to

30.0 mg/L (ppm) range. The method is an adaptation of the Standard Method for the
Examination of Water and Wastewater.
The reagents are in

powder and liquid form and are supplied in packets and bottles. The amount

of reagent is precisely dosed to ensure the maximum repeatability.

degree Celsius
degree Fahrenheit
milligrams per liter. mg/L is equivalent to ppm (parts per million)
milliliter
millivolts

°C

::

°F

:

mg/L

:

mL

:

mV:

Range

0.0 to 30.0 mg/L

Resolution

0.1 mg/L

Accuracy

±1.0 mg/L ±4% of reading @ 25°C

Typical EMC Deviation

±0.2 mg/L

Light Source

Tungsten lamp

Light Detector

Silicon Photocell with narrow band interference filter @ 525 nm

Method

Amino Acid Method, adapted from Standard Method for the Examination
of Water and Wastewater. The reaction between phosphate and the
reagents causes a blue tint in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in

measurement mode;

after 1 hour of non-use in

calibration mode;

with last reading reminder.

Dimensions

192 x 102 x 67 mm (7.6 x 4 x 2.6")

Weight

290 g (10 oz.).

REQUIRED REAGENTS
Code

Description

Quantity/test

HI 93717A-0

Phosphate Reagent A (Molybdate)

10 drops

HI 93717B-0

Phosphate Reagent B (Amino Acid)

1 packet

Precision

Precision

Precision

Precision

Precision is how closely repeated measurements
agree with each other. Precision is usually
expressed as standard deviation (SD).
Accuracy

Accuracy

Accuracy

Accuracy

Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of 12.0
mg/L phosphate and a representative lot of
reagents, an operator obtained with a single
instrument a standard deviation of 0.5 mg/L.