Hanna Instruments HI 96717C User Manual

Page 7

Advertising
background image

12

13

• Proper use of dropper:

(a) to get good reproducible results, tap the dropper on the table for several times and wipe

the outside of the dropper with a cloth.

(b) always keep the dropper bottle in a vertical position while dosing the reagent.

• Proper use of the powder reagent packet:

(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet.

• Shaking the cuvette can generate bubbles in the sample, causing higher readings. To obtain

accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette.

• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row, but it is recommended to take a new zero

reading for each sample and to use the same cuvette for zeroing and measurement.

• After the reading it is important to discard immediately the sample, otherwise the glass

might become permanently stained.

• All the reaction times reported in this manual are referred to 20°C (68°F). As a general rule

of thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a measurement follow the validation procedure to

be sure that the instrument is properly calibrated. If necessary, calibrate the instrument.

(a)

(b)

MEASUREMENT PROCEDURE

STARTUP

Prepare the instrument for measurement as follows:
• Unpack the instrument by removing the dust protection sleeve from the instrument cuvette

holder.

• Place the battery in the instrument as described in the “BATTERY REPLACEMENT” chapter.
• Place the instrument on a flat table.
• Do not place the instrument under direct sun light.

To compensate the meter for the sample turbidity or color, the measurement takes place in two
phases. First, the meter is zeroed using the unreacted sample. After the reagents are added the
reacted sample is measured.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.
The blinking “ZERO” indicates that the
instrument needs to be zeroed first.

• Fill the cuvette with 10 mL of unreacted

sample, up to the mark, and replace the
cap.

• Place the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Turn the meter on by pressing ON/OFF

ON/OFF

ON/OFF

ON/OFF

ON/OFF.

The display briefly shows all tags on.

Advertising