BROMINE MEASUREMENT

BROMINE MEASUREMENT

BROMINE MEASUREMENT

BROMINE MEASUREMENT

BROMINE MEASUREMENT

• Fill the cuvet up to the mark with 10 mL of

unreacted sample and replace the cap.

• Place the cuvet into the holder and

ensure that the notch on the cap is
positioned securely into the groove.

• Press ZERO and "SIP" will appear on the display.

• Wait for a few seconds and the display will

show "-0.0-". Now the meter is zeroed and
ready for measurement.

• Remove the cuvet and add the

content of one packet of HI 93716
reagent. Replace the cap and shake
gently.

• Reinsert the cuvet into the instrument.

• Wait for 2 minutes and 30 seconds and press READ. "SIP" will

appear during measurement.

• The instrument directly displays concentration in mg/L of bromine

on the Liquid Crystal Display.