Hanna Instruments HI 93710 User Manual

Page 35

Advertising
background image

35

• Reinsert the cuvet into the instrument.

• Wait for approximately 30 seconds and then press READ. The

display will show “SIP” during measurement.

• The instrument directly displays concentration in mg/L of free

chlorine on the Liquid Crystal Display.

Note: Free and total chlorine have to be measured separately with

fresh unreacted samples following the above procedure if both
values are requested.

INTERFERENCES

Interference may be caused by:
Bromine
Iodine
Fluorine
Ozone
Oxidized manganese and Chromium
In case of water with hardness greater than 500 mg/L CaCO

3

, shake

the sample for approximately 1 minute after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO

3

or

acidity greater than 150 mg/L CaCO

3

, the color of the sample could

disappear or develop only partially. To resolve this, neutralize the
sample with diluted HCl or NaOH.

Free Chlorine