General tips for an accurate measurement – Hanna Instruments HI 96726 User Manual

Page 11

Advertising
background image

11

The instructions listed below should be carefully followed during testing to ensure best accuracy.
• Color or suspended matter in large amounts may cause

interference, therefore these should be removed by
treatment with active carbon and by prior filtration.

• For a correct filling of the cuvette: the liquid in the

cuvette forms a concavity on the top; the bottom of this
concavity must be at the same level of the 10 mL mark.

• Proper use of the powder reagent packet:

(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet.

• It is important that the sample does not contain any debris. This would corrupt the reading.
• Each time the cuvette is used, the cap must be tightened to the same degree.

GENERAL TIPS FOR AN ACCURATE MEASUREMENT

Battery low:

The battery must be replaced soon.

Dead battery:

This indicates that the battery is dead and must

be replaced. Once this indication is displayed, the meter will
lock up. Change the battery and restart the meter.