Hanna Instruments HI 96726 User Manual

Page 14

Advertising
background image

14

• Remove the cuvette.

• Add the content of one packet of

HI 93726-0

reagent. Replace the cap

and shake gently.

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 1 minute then just
press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the instrument

directly displays concentration in g/L of
nickel on the LCD.

INTERFERENCE
Interference may be caused by copper.

Advertising