• Remove the cuvette.
• Add the content of one packet of
HI 93726-0
reagent. Replace the cap
and shake gently.
• Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
• Press and hold READ/ /TIMER
for three
seconds.
The display will show the
countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 1 minute then just
press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
• At the end of measurement, the instrument
directly displays concentration in g/L of
nickel on the LCD.
INTERFERENCE
Interference may be caused by copper.