2 switch the instrument on, Instrument on, On, instrument – Heath Consultants Personal Surveyor (PS200 Series) User Manual
Page 26
 
2-2
USER HANDBOOK
•
The instrument is within the calibration period you 
have decided is necessary for your application. 
•
If oxygen sensor is fitted, check oxygen readings to 
ensure correct operation. The oxygen sensor 
responds to the user breathing on the instrument 
front face (sensor area) by displaying a decreased 
value, i.e. below 20.9%. 
Switch the instrument OFF, in fresh air, after use.
Additional:
•
Perform regular leak check on pumped instruments, by 
placing thumb over sample inlet nozzle and making sure 
that instrument displays ‘FLOW FAULT’. 
•
Perform regular bump tests using either the Auto Bump 
/ Calibration Station or by performing manual bump tests. 
2.2 SWITCH THE INSTRUMENT ON
Press and hold the Right Hand (RH) button
for one
second to switch the instrument ON. Refer to Fig. 2-1.
The instrument begins its warm-up routine, which lasts 
approximately 30 seconds. During the warm-up, a countdown 
timer appears in the top (RH) corner of the display. 
Note: The display backlight illuminates green and remains 
ON during warm-up. When warm-up is complete, the screen 
light automatically switches off.