MKIII SERVICE MANUAL

30


Exit out of the Brewcmd program by typing quit

and the computer will have returned to the

dos prompt.
Run the BREWER operating program and then ensure the time and date is correct.


8.3

NEW CONFIGURATION UPLOAD


If hardware changes have been made such as removing the zenith drive gear or adjusting the
micrometer drive sensor mask, then correct values for zero offset and other parameters must be
uploaded to the instrument for it to operate correctly. The procedure to upload the new
configuration file is listed below:
The configuration file that describes the characteristics of the instrument is named BREWnnn.cfg.
The nnn will be the instrument number. This file is an ASCII text file that describes all the technical
parameters of this particular instrument. This very same file should be uploaded to any
replacement BREWER Main Electronics board following the physical installation of the board. A
spare BREWER Main Electronics board will normally be sent with the latest configuration file stored
at KIPP & ZONEN. If the customer has changed any parameters in the field, it will not be
configured into that board and the customer will have to upload the latest file as described below:

Change to the directory c:\bdata\nnn by typing the command

cd \bdata\nnn

Run Brewcmd.exe by typing brewcmd

[using cosmac mode at 1200 baud, on com 1: tracing is off ]should be
displayed
Type loadmode

to change modes.

Waiting for the mode change to complete and when complete, the display will display
[using load mode at 1200 baud, on 1: tracing is off ]
Go into Opmode by typing opmode 9600

Trying Operating mode at 9600 baud is displayed until the display will display [using
operating mode at 9600 baud; on port 1; tracing is off ]

Send the configuration file to the microprocessor by typing putcfg brewnnn.cfg

 

(nnn is the

instrument number)
Wait approximately one minute to transfer
Type readlog

to clear the log buffer

Type useconfig

motors should now reset

When complete readlog

to see if errors were encountered. Contact KIPP & ZONEN if the log

displays a problem.
Type save

a number should appear. This tells you how many configuration writes you have left.

If the number is ‘1’ then the next time a configuration is sent, the top firmware must be written to
Flash memory and the configuration area is cleared to allow another 4 blocks of configuration files.
Go to Load mode by typing loadmode

Waiting for mode change to complete and
Trying load mode at 9600 baud is displayed until the display will display [using load
mode at 9600 baud, on com 1: tracing is off ]


Go into Cosmac mode by typing cosmacmode

Reading Cosmac mode reset message
Waiting for mode change to complete
Trying Cosmac mode at 1200 baud
Motors Initializing