Timer operation, Interval timer operation, Stopwatch timer operation – Labconco Biological Safety Cabinets 30268 Series User Manual

Page 41: If an airflow alert activates, Resetting the airflow alert system

Advertising
background image

Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

35

Original instructions

Timer Operation

NOTE: The timer button allows activation of an interval (countdown) or
stopwatch (elapsed) timer. The timers cannot be operated
simultaneously.

To access the main timer menu, press the

Timer

button anytime during

normal operation. The main timer menu is shown on the LCD display. Use
the

and

buttons to highlight the

Interval

or

Stopwatch

Timer. Press the

OK

button to select the highlighted timer function.

Interval Timer Operation

1. When selected, the Interval Timer menu is displayed on the LCD.

The timer defaults to 0:00:00 (hours:minutes:seconds).

2. Press and hold the

or

buttons to increase or decrease the timer

interval.

3. When the proper interval is entered on the display, press the

OK

button to start the timer.

4. When the timer reaches 0:00:00, an audible alarm will sound, and the

timer will reset itself and repeat the countdown.

5. Press the

OK

button to pause the timer.

6. Press the

Menu

button to clear the interval timer and return to the

main timer menu.

Stopwatch Timer Operation

1. When selected, the Stopwatch Timer menu is displayed on the LCD.

The timer defaults to 0:00:00.

2. Press the

OK

button to start the timer.

3. Press the

OK

button again to zero the timer.

4. Press the

Menu

button to clear the stopwatch timer and return to the

main timer menu.

If An Airflow Alert Activates

The most common causes of an Airflow Alert are:

• Blockage of the inlet grilles or exhaust outlet.

• Removal of the work surface or grille during operation.

Resetting the Airflow Alert System

The Airflow Alert automatically resets to normal operation once the motor
speed has stabilized.

Advertising