Chapter 6: Maintaining The Cabinet

Product Service 1-800-522-7658

44

Original instructions

4. Unplug the cabinet or turn off the System Reset Switch, located on the

top of the cabinet.

5. Remove the UV lamp by rotating it 90 degrees and lifting it straight up

and out of its sockets.

6. Install new lamp by reversing the removal procedure.

Resetting a Circuit Breaker:

To reset any of the circuit breakers located on the left side of the electronics
module, depress the white button until it sets.


Circuit Breakers- The
front breaker protects
the electrical outlets,
the rear breaker
protects the motor and
lights.










Storage

If the biosafety cabinet is to be left unused for more than one month, it should
be prepared for storage.

Note: The cabinet should not be stored in areas of excess humidity or
temperature extremes. If the cabinet is moved during storage, it must be
recertified before use.
1. Close the sash completely and seal the bottom edge and the exhaust outlet

with plastic sheeting.

2. Unplug the cabinet.

3. Ensure that the cabinet will not be moved or disturbed while being stored.

Figure 6-5