Ppendix, Uick, Hart – Labconco Class I Enclosure Models 3980423 User Manual
Page 73: Iltered, Nclosures
Product Service 1-800-522-7658
69
A
PPENDIX
D
Q
UICK
C
HART
F
OR
T
HE
F
ILTERED
E
NCLOSURES
To determine the actual face velocity at the sash opening, airflow velocity readings
will need to be taken. This should be done across the sash opening of the
enclosure in accordance with the Industrial Ventilation Manual. (See Appendix E)
The “average face velocity” is achieved by taking readings in two rows across the
enclosure with the readings 6" from the ends and evenly spaced every 12"; the first
row is 3" down from the upper sash foil and the second row is 3" up from the work
surface. A total of four readings will be taken for the 2' enclosure, six readings
taken for the 3' enclosure and eight readings taken for the 4' enclosure and then
averaged.
Model Size
2'
3'
4'
Sash Height from Work Surface (inches)
9.44
9.44
9.44
Total Open Area with Bypass (sq. ft.)
1.45
2.20
2.95
Exhaust Airflow Volume at 90fpm (CFM)
130
200
265
Alarm Airflow Volume at 60 fpm (CFM)
85
130
175
Initial Mag. Gauge Pressure Reading (inches H
2
O)
0.20 to .045
0.15 to 0.40
0.10 to 0.35
Number of Laskin Nozzles needed at 10 psig
1
1
1
Exhaust HEPA Filter w/ Gasket Dimensions (inches)
18x18x3.31
30x18x3.31
42x18x3.31
*Motor Horsepower (HP)
0.13
0.13
0.30
*Capacitor (MFD, Volts)
10MFD, 115V
(2) 4MFD, 230V
10MFD, 115V
(2) 4MFD, 230V
15MFD, 115V
4MFD, 230V
Fluorescent Lights
F15T8-SP35
F17T8
F25T8
UV Light on Class I only
G15T8
G15T8
G30T8