Tap water supply, Distillation manifold – Labconco Open Combination (Macro) Kjeldahl Systems 2123Series User Manual

Page 12

Advertising
background image

Chapter 2: Installation

Product Service 1-800-522-7658

8

Tap Water Supply

NOTE: ALL PLUMBING CONNECTIONS AND COMPONENTS MUST
BE FREE OF FOREIGN MATERIAL BEFORE FINAL CONNECTIONS
ARE MADE
.

Distillation Manifold

The Open Combination Macro-Kjeldahl Systems feature an upper distillation rack
with individual condensation columns, these are connected together by a common
manifold (6 columns per manifold). The tap water inlet and outlet connections are
3/4" NPT and the supply line should be minimum 1/2" ID for proper water flow
and heat removal capacity through the condensation columns (6-8 gallons/23-30
liters per minute). Open Combination Macro-Kjeldahl Systems with 12 or 18
heaters per rack will include two or three, 6 column manifolds which are
connected together in series.

The plumbing line for the outlet of the distillation manifold must be free of back
pressure to avoid flow restriction of the cooling water.

Water discharged from the distillation manifold is not acid contaminated and
standard drain lines may be used for its removal.















Distillation Manifold Tap Water Connections

Figure 2-3

Advertising