Chapter 2: Installation

Product Service 1-800-522-7658

9

Acid Fume Blower Exhaust System

The exhaust duct connection on top of the damper is sized for use with 6" nominal
(6-5/8" O.D.) vent duct. The exhaust piping assembly is supplied with a short
piece of 6" PVC pipe and a flexible coupling with clamps to connect to the
dedicated duct work exhaust. The flexible coupling must be sealed air-tight to the
exhaust duct work.

All exhaust duct work must be acid-resistant and sealed air-tight. Use only acid-
resistant sealant on duct work joints.



















The bottom of the piping assembly is fitted with a condensate drain and corrosion
resistant tubing. This condensate (identified above and on the next page) can be
collected in a closed acid proof container, which should be emptied regularly or it
may be plumbed to an acid resistant drain line. The condensate is concentrated
acid and care must be exercised when handling it.

Blower Airflow Exhaust Volume Requirement &
Velocity Test

Cubic feet per minute = 25
Average airflow velocity = 117 feet per minute

The Kjeldahl’s airflow volume test is based on the "Pitot Traverse in Inlet Duct"
method described in the ANSI/ASHRAE 51-1999 standard. Duct velocity
measurements are taken at 48" from the flexible coupling using a 6 point duct
traverse pattern (Industrial Ventilation 24th Edition p. 9-10).

Figure 2-6

Drain Line

Damper

Flexible 6”
Coupling

Piping
Assembly