Ppendix, Pecifications – Labconco Purifie Non-Ventilated PCR & Tissue Culture Enclosures 3952420 User Manual
Page 27
![background image](/manuals/738666/27/background.png)
Product Service 1-800-522-7658
23
A
PPENDIX
C
S
PECIFICATIONS
This Appendix contains technical information about all the Purifier Passive PCR
Enclosures and Purifier Tissue Culture Enclosures including electrical specifications
and environmental conditions.
Electrical Specifications
• 10 Amps, 115V, 60 Hz
• 5 Amps, 230V, 50/60 Hz
• The auxiliary outlet load is 8 Amps on 115V and 4 Amps on 230V
Environmental Conditions
• Indoor use only.
• Maximum altitude: 6562 feet (2000 meters).
• Ambient temperature range: 41° to 104°F (5° to 40°C).
• Maximum relative humidity: 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
• Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
• Transient over-voltages according to Installation Categories II (Over-
voltage Categories per IEC 1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for 115V models and 2500V for
230V models are allowed.
• Used in an environment of Pollution degrees 2 (i.e., where normally only
non-conductive atmospheres are present). Occasionally, however, a
temporary conductivity caused by condensation must be expected, in
accordance with IEC 664.