This Appendix contains technical information about all the Purifier Passive PCR
Enclosures and Purifier Tissue Culture Enclosures including electrical specifications
and environmental conditions.

Electrical Specifications
• 10 Amps, 115V, 60 Hz

• 5 Amps, 230V, 50/60 Hz

• The auxiliary outlet load is 8 Amps on 115V and 4 Amps on 230V


Environmental Conditions
• Indoor use only.

• Maximum altitude: 6562 feet (2000 meters).

• Ambient temperature range: 41° to 104°F (5° to 40°C).

• Maximum relative humidity: 80% for temperatures up to 88°F (31°C),

decreasing linearly to 50% relative humidity at 104°F (40°C).

• Main supply voltage fluctuations not to exceed ±10% of the nominal

voltage.

• Transient over-voltages according to Installation Categories II (Over-

voltage Categories per IEC 1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for 115V models and 2500V for
230V models are allowed.

• Used in an environment of Pollution degrees 2 (i.e., where normally only

non-conductive atmospheres are present). Occasionally, however, a
temporary conductivity caused by condensation must be expected, in
accordance with IEC 664.