Labconco CentriVap Mobile Systems 7812014 User Manual

Page 7

Advertising
background image

Product Service 1-800-522-7658

3










C

C

h

h

a

a

p

p

t

t

e

e

r

r

2

2

:

:

P

P

r

r

e

e

r

r

e

e

q

q

u

u

i

i

s

s

i

i

t

t

e

e

s

s

Before you install your CentriVap, you need to prepare your site for installation.
Carefully examine the location where you intend to install your CentriVap. You
must be certain that the area is level and of solid construction. In addition, an
exhaust means must be provided. An electrical source must be located near the
installation site.

Carefully read this chapter to learn:

• The electrical supply requirements.

• The exhaust requirements.

• The vacuum pump requirements.

Refer to Appendix C: CentriVap Specifications for complete CentriVap electrical
and environmental conditions, specifications and requirements.

Electrical Requirements

The CentriVap Mobile System requires a dedicated grounded electrical outlet.
This outlet requires a 20 Amp circuit breaker or fuse for models rated at 115V
(60 Hz). A 10 Amp circuit breaker or fuse is required for models rated at 230V
(50/60 Hz). 115V models are equipped with a 20 Amp NEMA 5-20P plug. The
power cord on 230V models is equipped with a NEMA 6-15P plug. If this does
not match with the available receptacle, remove this plug and replace it with an
approved plug of the suitable style.

It is recommended that an emergency switch for disconnecting the mains in the
case of a malfunction be located remote from the CentriVap, preferably outside
the room in which the CentriVap is housed, or adjacent to the exit from that room.

Advertising