Labconco Guardian 1000 Digital Airflow Monitor User Manual

Page 18

Advertising
background image

2

FILE NUMBER

AFA 1000 / 1 -- ALARMS -- ( Velocity )

TYPE

CUSTOMER



FACTORY DEFAULTS

The AFA1000 alarm has been factory set to the listed parameter settings. To change

any parameter setting press and hold the enter button until the "Main Menu" screen is

displayed, select "Configure" to view the parameters and use the +/- and Enter buttons

to change the settings.

FOR FULL INSTALLATION / CALIBRATION / CONFIGURATION DETAILS PLEASE

REFER TO THE DATA BOOKLET PROVIDED.

Configuration Menu: -

Menu

Calibration Config

Units

Default Settings

M/SEC

ON DISPLAY

UNITS

FPM

FPM

ON

LOW AIR ALARM

FPM

60

OFF

ON LOW

AIR

CUTOFF

ON

ON

ON CUTOFF

SPEED

FPM

30

WARNING AIR ALARM

FPM

63

ON

ALARM TO WARNING AIR HYSTERESIS

FPM

3

YES

ON

HIGH AIR ALARM

NO

OFF

ON

HIGH AIR ALARM

FPM

295

ON

LOWER AIR SAMPLE FLUCTUATIONS

%

3

ON

HIGHER AIR SAMPLE FLUCTUATIONS

%

3

ON LOWER/HIGHER AIR SAMPLE DIFFERENCE

FPM

20

ON

WARN TO ALARM AIR TIME

SECONDS

5

ON

ALARM TO WARN AIR TIME

SECONDS

1

YES

ON

ON SHOW

AIR

FLOW

NO

YES

ON

SHOW TIME LINE

NO = SHOW BAR GRAPH

OFF

ENABLED

ENABLED

ON AUDIBLE

ALARM

NOT ENABLED

ON SENSOR

DIFFERENCE

%

10

Configuration Menu

NOT ACTIVE

ACTIVE ON CLOSING CONTACT

ACTIVE ON CLOSING CONTACT

ON INPUT

1

ACTIVE ON OPENING CONTACT

NONE

ALARM DISABLE

NIGHT SET-BACK

NIGHT SET-BACK

EXTERNAL ALARM

SASH HIGH

ON

INPUT 1 - FUNCTION

HIGH/LOW