2 alarm configuration / calibration, 3 start up – Labconco Guardian 1000 Digital Airflow Monitor User Manual

Page 7

Advertising
background image

7

1.2 Alarm Configuration / Calibration



The alarm can be configured via a Laptop or PC using a variety of ‘set up ’ programs each
designed for a particular application with a combination of inputs , outputs and push buttons.
This configuration can be changed via the alarm key pad using the menu system if required or
re-configured by re-connection of the laptop or PC.

This allows the fume hood manufacturer to stock standard units and configure the alarms
to suit the application.

The configuration of the various functions and the calibration of the alarm face velocity display is
menu driven. Access to the menu will be via password ( 4 digit number ) and will be two level.
The first level will be for calibration of the unit and the second level will be for ‘engineers’ to set up
the configuration of the alarm.

NOTE:- If you enter the Calibration or Configure Menu by accident :-
press the + & - buttons at the same time to escape back to the Main Menu


The menus and sub–menus are in ‘ plain language ’ and incorporate brief instructions where
appropriate.

See menu operation document



1.3 Start up


When unit is powered up the following sequence of events occur :-


1. The 12V DC power is applied to the airflow sensor and a delay on timer ( 30 sec ) is initiated.
2. The alarm then performs a self test on the display and all indicators etc ( approx 5 sec )

3. The display show a ‘ Welcome note ’ – with the fume cupboard manufacturer company name

( if configured ) for the rest of the initial 30 sec delay time. During this time the airflow sensor
is stabilising


4. During the whole of the 30 sec period all alarms and relay outputs are inhibited.

5. At the end of the 30 sec delay the unit performs one of two options :-

a. If the alarm calibration has been previously completed – the unit goes to
normal operating mode ( Run )

b. If the unit has not been calibrated the unit displays
‘ Unit requires Set up -- press Enter to access Set up menu ’

The set up menu allows calibration or configuration via the password protection



During the set-up all alarms and output relays are inhibited.