To install
- Unpack new lung/stomach (Cat. No. 08 20 25).

- Connect stomach tube to oesophagus end of head/

airway assembly.

- Place new airway/stomach with the lung connectors into

the two holes in the chest plate.

- Connect the two lung tubes to the two lung connectors

in chest plate.

- Mount the stomach pressure valve band over the

two tabs. Put chest plate in place over the chest
compression slide.

- Put chest skin back on. Make sure the cables are

placed inside the chest skin so that they do not restrict
chest movement.

Cleaning upper airways

To remove
- Open up manikin and disconnect airway/stomach as

described in "To remove" in section "Replacing the
lungs/stomach".

- Press down tab on inner

neck connector and pull
head away from body.

- Remove face skin as described in "To remove"

in section "Replacing or cleaning the face skin".

-

Unsnap the two lung
connectors.

-

Unsnap the silicone stomach
tube from the oesophagus
end of the head/
airway assembly.

-

Discard lung/stomach.

To clean
- Open and empty the cleaning kit basin.

- Place the manikin head face

down in the cleaning basin.

- Connect the two lung tube

connectors to the double
Y-piece in the tube system.

- Connect the oesophagus

to the single connection
in the tube system.

- Insert the cleaning syringe into the valve opening

and place the free end of the tube system into the basin.

- Fill the basin with liquid to just

underneath the edge. Circulate
liquid through the manikin airways
by pumping the syringe plunger.

- Use this setup to go through

the following steps:

1: Use a Disinfectant solution*. Airways should remain

completely filled for at least 10 minutes.

2: Use fresh water to remove disinfectant solution.

After each step lift the head clear of the liquid

to allow drainage.

3: Allow to dry completely before reconnecting head

to manikin.

* Virkon can be used. Depending on local procedures you may prefer to
use a sodium hypochlorite solution freshly prepared for single class use.
This solution should have at least 500 ppm free available chlorine, i.e. 22 ml of 2.5 -
5.0 % household bleach per 500 ml of water. Local cleaning and handling procedures
must be followed.