Upright, Testing before use, Reprocessing instructions – Laerdal LGH Upright Bag-Mask User Manual

Page 2: Newborn bag - mask

Advertising
background image

3.Testing before use

Squeeze the bag. Check that the yellow valve

opens and closes with every squeeze.

Keep the mask sealed against your hand.

Release the squeezed bag. Check that the

bag re-expands without resistance.

Keep the mask sealed against your hand.

Press the pressure release valve down.

Squeeze the bag and check that there is

no leakage.

Seal the mask against your hand. Squeeze

the bag forcefully. Check that air is

released from the pressure release valve.

If any of the above tests fail: Dismantle Upright, inspect the components, reassemble Upright and repeat the complete “Testing before use” procedure (Section 3).

60 minutes

10 minutes

100 °C

Disinfect

Always dismantle

Upright before

cleaning.

1

2

3

4

Wash all parts in a

clean tray with clean

water and mild soap.

Use a scrub or brush

to remove any soil.

Rinse parts in clean

water to remove all

soil and soap.

Repeat above steps

until parts are clean.

Dry all parts.

Visually inspect

each part for

damage and

cleanliness /

mineral deposits.

Remove damaged

or unclean parts

from service.

Reassemble Upright.

Test usings steps

shown below.

OR

OR

2. Cleaning and Disinfection procedure

1. Product overview

1. Dismantle

2. Clean

3. Disinfect by one of the methods

4. Dry and Inspect

5. Assemble and Test

1

2

3

5

6

Pull Out

Unscrew

Lift Out

Unscrew

Pull Out

To dismantle, follow steps 1- 6. To reassemble, follow the steps in reverse.

Do not disassemble parts beyond the steps shown. For proper function, make sure to only use parts marked LAERDAL (locations as shown above).

Upright

NEWBORN BAG - MASK

4v

Pull forcefully on tab to release

Chemical

Activated glutaraldehyde solution:

1

2

3

4

Autoclaving

Sterilize by steam

autoclaving

at 136 °C and

2.0 kg/cm

2

for

10-20 min.

Boiling*

Boil all parts in clean

water for 10 min.

10-20 minutes

Steam 136 °C

Reprocessing instructions

Page 2

Disinfect all parts to the dis-

infectant manufacturer’s

recommended procedure.

Immerse for 60 minutes.

Remove parts using aseptic

technique.

Rinse 3 times, for 1 minute using

fresh clean water each time.

1

2

3

1. Lip valve function

2.

Inlet valve opening

4. Product sealing

3.

Pressure release valve

“LAERDAL”

“LAERDAL”

+ rinse

in water

3 x 1 minute

*Validated at approximately

sea-level pressure

Check that the
silicone bag's lip

is not twisted or
tucked under.

4