Lamar Technologies Activator 282 User Manual

Page 11

Advertising
background image

10

Verification of output is the only check that is required on an annual basis.

Voltage test points are provided on the unit and verification can be accomplished

with a calibrated voltmeter such as the Fluke 77 / AN.

If the Activator 282 supplies the correct voltage, the required charge current will

automatically follow, and charge amperes are a consequence of the charger and

battery voltage. Charge current from the Activator 282 is not adjustable, and is

automatically limited to the capacity of the Activator 282: 10 Amp pulses for a low

battery voltage, and 2 ~ 5 amp steady state).

6.2

OUTPUT VOLTAGE VERIFICATION AND CALIBRATION PROCEDURE

Tools and materials:

1)

Activator 282 to be verified.

2)

Calibrated voltmeter with 0.080” (2mm) tip probes.

3)

Fine-tipped (1/16”) flat blade screwdriver

Procedure:

a) It does not matter if the Activator 282 is connected to a battery. It does not

matter if the Activator 282 is set for a Lead Acid or NiCad battery.

b) Connect a calibrated voltmeter to the red (+) and black (-) tip jacks on the

Activator 282.

c) Turn on the Activator 282 and observe both its displayed voltage and the

voltmeter voltage. This should be between 28 and 35 Volts.

d) The Activator 282 must agree with the voltmeter within ± 0.3 Volts. If not,

locate the adjustment hole on top of the Activator 282, at the fan outlet and

near the most significant digit of the voltmeter display. Adjust the control with

a fine-tipped blade screwdriver to bring the indicated voltage into agreement

with the external calibrated voltmeter.

6.3

CHECKING THE CURRENT

Tools and materials:

Advertising