5 Calibration - Using an External Calibration Weight

Calibration is required when the weighing machine is initially installed or if the Balance is moved to a
substantial distance from the original location. 30 minutes of warming up of the Balance is always
needed before weighing and calibration.

Turn the Balance on and when display shows

“--------”, press ZERO key

The display will show “CAL”

Press ZERO key again, the display will show “X000.00”, “X” is the flashing digit

Press UNIT key, the flashing digit will move to right

Press PCS key to increase the value of the flashin

g digit(“X” is the calibration weight can be set

according to users’ requirements, we recommend a minimum weight of at least 50% of the
Balance capacity)

Press ZERO key, display will show “CAL” and then series of the digit, wait for stable indication
shows, press UNIT key , display will show --------

, then “X000.00”

Place the corresponding test weight on the center of the pan

The display backlight lits and returns to weighing mode

Now the calibration is completed

Turn off the power, then turn on the power again, place a known weight on the platform to verify if
it is accurate.

If not, repeat above steps.