LW Measurements KHR 502 User Manual

Page 21

Advertising
background image

21

5 Calibration - Using an External
Calibration Weight

Calibration may be required when the weighing machine is initially installed or

if the weighing machine is moved a substantial distance from the original

location. Thirty (30) minutes of warming up of the machine is always needed

before initial calibration.

Calibration procedure:

Before entering the Calibration Mode, press [ON / OFF] to power on
the scale for 1 minute.

Press [ON / OFF] to turn the power off.

Press and hold the [UNIT] key and at the same time press the [ON /
OFF] key. The display will show CALu=. The scale is now in
Calibration Mode.

Press the [UNIT] key to toggle between setting units (kg, g, lb, oz).

Press [ZERO] to initiate the calibration process.

Enter the weight of your calibration weight by using the [UNIT] key to
toggle between digits, and use the [TARE] key to increase each digit
until the display flashes your test weight amount.

Press the [ZERO] key. The display will now show CAL then AD value.

Press the [UNIT] key. The display will show the flashing calibration
weight.

Place your test weight on the platter. Wait for the Stable Indicator to
be displayed. Press the [UNIT] key. The display will return to the AD
value