RVC 2-33 CD

&

Maintenance and Service

Version 01/2009, Rev. 1.21 of 25/11/2013 - sb

55

Translation of the original operating manual

8.1.2 Rotor Chamber

If the rotor chamber is contaminated or if glass is broken, the rotor
chamber, the rotor shaft and the rotor must be cleaned immediately in
order to prevent corrosion.

Cleaning of the rotor chamber:

1 Rotor chamber
2 Cap
3 Rotor bearing

Fig. 8.1: Rotor chamber with rotor shaft

Open the lid.
Disconnect the power cord from the wall outlet or instrument

receptacle before cleaning.

Hot surface!
Let the rotor chamber, the rotor shaft and the lid cool down to room
temperature!

Pull off the cap.

1 Allen screw

Fig. 8.2: Removing the allen screw

Remove the allen screw with the supplied allen wrench (size 5).

1

1

2

3