3 autoclaving, 4 service – Martin Christ RVC 2-33 CDplus User Manual

Page 59

Advertising
background image

RVC 2-33 CD

&

Maintenance and Service

Version 01/2009, Rev. 1.21 of 25/11/2013 - sb

59

Translation of the original operating manual

8.3 Autoclaving

The service life of the accessories essentially depends on the
frequency of autoclaving and use.

Remove the existing O-rings so that the trapezoid notch can be

reached by the hot steam. This prevents the formation of humidity
in the notch..

Replace the accessories immediately when the parts show

changes in color or structure or in the occurrence of leaks etc.

Accessories

max. temp.

°C

min. time

min

max. time

min

Cycles

(max.)

Glass tubes

134-138

3

40

-

Polycarbonate tubes

115-118

30

40

20

Polypropylene tubes

115-118

30

40

30

Teflon tubes

134-138

3

5

100

Aluminium rotor

134-138

3

20

-

Aluminium rotor accessories

134-138

3

20

-

8.4 Service

In the event of service work that requires the removal of the panels,
there is a risk of electric shock or mechanical injury. Only qualified
specialist personnel is authorised to perform this service work.

The rotational vacuum concentrator is subject to high mechanical
stress. In order to be able to withstand this high level of stress, high-
quality components were used during the production of the rotational
vacuum concentrators. Nevertheless, wear cannot be excluded and it
may not be visible from the outside.
This is why we recommend having the rotational vacuum
concentrator checked by the manufacturer during an inspection once
per year.

Information and appointments:
in Germany:

use the service request form at

www.martinchrist.de

[Service Area]

or contact

Martin Christ Gefriertrocknungsanlagen GmbH

An der Unteren Söse 50
37520 Osterode (Germany)
Tel. +49 (0) 55 22 / 50 07-8425
Fax +49 (0) 55 22 / 50 07-9425
E-mail: [email protected]


in other countries:

Advertising