Proximate Interface

w w w . p a r r i n s t . c o m

5

Explanation of Symbols

I

On position

O

Off Position

~

Alternating Current (AC)

This

CAUTION

symbol may be present on the Product Instrumenta-

tion and literature. If present on the product, the user must consult
the appropriate part of the accompanying product literature for more
information.

ATTENTION

, Electrostatic Discharge (ESD) hazards. Observe precau-

tions for handling electrostatic sensitive devices.

Protective Earth (PE) terminal

. Provided for connection of the Protec-

tive Earth (green or green/yellow) supply system conductor.

Chassis Ground

. Identifi es a connection to the chassis or frame of the

equipment shall be bonded to Protective Earth at the source of supply
in accordance with national and local electrical code requirements.

Earth Ground

. Functional earth connection. Note: This connection shall

be bonded to Protective earth at the source of supply in accordance
with national and local electrical code requirements.

General Specifi cations

Electrical Ratings
115VAC, 2.0 Amps. 50/60 Hz
230VAC, 2.0 Amps, 50/60 Hz

Before connecting the 6750 Proximate Interface

to an electrical outlet, the user must be certain that
the electrical outlet has an earth ground connection
and that the line, load and other characteristics of
the installation do not exceed the following limits:

Voltage: Fluctuations in the line voltage should not
exceed 10% of the rated nominal voltage shown on
the data plate.
Frequency: 6750 Proximate Interface units can be
operated from either a 50 or 60 Hertz power supply
without affecting their operation or calibration.
Current: The total current drawn should not exceed
the rating shown on the data plate on the unit by
more than 10 percent.

Environmental Conditions

Operating:

15 ºC to 40 ºC; maximum relative humidity of
80% non-condensing.

Installation Category II (overvoltage) in accor-
dance with IEC 664.
Pollution degree 2 in accordance with IEC 664.
Altitude Limit: 2,000 meters.

Storage:

-25 ºC and 65 ºC; 10% to 85% relative humidity.

Cleaning & Maintenance

Periodic cleaning may be performed on the

exterior surfaces of the instrument with a lightly
dampened cloth containing mild soap solution. All
power should be disconnected when cleaning the
instrument. There are no user serviceable parts
inside the product other than what is specifi cally
called out and discussed in this manual. Advanced
troubleshooting instructions beyond the scope of
this manual can be obtained by calling Parr Instru-
ment Company in order to determine which part(s)
may be replaced or serviced.

Fuses

Lines protective fuses rated: Fast-Act 15 Amp,

250VAC (Parr number 139E23).

The replacement of protective fuses for the 6750

Proximate Interface should be performed by quali-
fi ed personnel.