Installation, Flat ptfe gasket, Self sealing o-ring closure – Parr Instrument Series 5500 User Manual
Page 6: Compact mini bench top reactors
 
Compact Mini Bench Top Reactors
P a r r I n s t r u m e n t C o m p a n y
6
INSTALLATION
Flat PTFE Gasket
The flat gasket is held in a recess in the vessel 
head and a machine pilot on the cylinder closes the 
recess to completely contain the gasket. The split 
ring closure used with this gasket has six cap screws 
which must be tightened to develop the loading on 
the gasket. 
The flat PTFE gasket can be used to operating 
temperatures as high as 350 ºC.
Self Sealing O-Ring Closure
The self sealing design features an O-ring retained 
in a groove on the vessel head. This design is self 
sealing and the split ring used with this sealing 
system does not require nor have the cap screws 
used with the flat gasket.
The maximum temperature of the vessels equipped 
with O-ring seals depends upon the material used 
for the O-ring. The most common material is a 
fluoroelastomer (FKM) which has a 225 ºC maximum 
operating temperature limit.
Moveable Head with PTFE Gasket
Moveable Head with O-ring Seal & Easy Close Split Ring