High temperature module - htm – Parr Instrument HTM User Manual

Page 6

Advertising
background image

- 6 -

HTM Installation

High Temperature Module - HTM


Keys command:

1. Press “SET” to select
2. Press return key move to next operation mode

3. Up/Down arrow keys to adjust value or select type


Main Screen: SV = 375

Press return key and release

Operation
Mode

Select
type/value

Comment

r-S Run Run/Stop

SP

0**

Decimal point position

AL1H 0

Upper-limit

alarm

1

LoC

OFF

Lock mode(lock all keys or only up/down arrow able to use)


**

If using a type-J or type-K thermocouple, SV = 0

If using an RTD, SV = 1


Press and hold down “SET” for 5-sec


Operation
Mode

Select
type/value

Comment

InPt J** Input

type

tPUn

C

Temperature unit

tP-H

800

Upper-limit temperature range

tP-L

0

Lower-limit temperature range

CTRL

ON/OFF

Control mode (ON/OFF, MANUAL, PID and PID

PROG)

S-HC Heat Heat/Cool

control

ALA1

2

Alarm operates when PV value is higher than SV value +

ALH setting

SALA OFF

System

alarm

CoSH

ON

Communication write function that able to use set point from

software

C-Sl ASCII Format

type

C-no 2

Communication

address

bPS

9600

Communication baud rate

LEn

7

Data length setting

PrtY

Even

Parity bit setting

StoP 1

Stop

bit

setting


**

If using a type-J thermocouple, InPt = J

If using a type-K thermocouple, InPt = K

If using an RTD, InPt = Pt