10. Connect the wires of the Cal 9500P

temperature controller as follows: (See Figure 4)

Wire:

Cal Pin #

Position:

Function

Black

7

Power

White

8

Power

Orange

24

High limit

Blue

23

High limit

Brown

21

SVM

Yellow

22

SVM

Red

19

SSR

Purple

20

SSR

T/C White

(Type J)

1

T/C

T/C Red

(Type J)

2

T/C

11. Remove the four wires connected to the

solid state relay on the back panel of the 4840

controller. (See Figure 5)

12. Remove the solid state relay from the back

of the 4840 controller and resistor if present.

Discard the old solid-state relay and resistor.

1

3. Install the provided solid-state relay (part #:

1119E) on the back panel of the 4840 controller.

(See Figure 6)

14. Attach the short black wire to position 1 of

the relay.

15. Attach the long black wire to position 2 of the

relay.

16. Attach the red wire to position 3 of the relay.

17. Attach the purple (or blue) wire to position 4

of the relay.

Back of Cal 9500P with wiring

Figure 4

Solid state relay for use with Watlow

(to be replaced)

Figure 5

Advertising