ADJUSTMENT INSTRUCTIONS CONTINUED
CORRECT
Above diagram shows example of CORRECT immersion on mattress overlay.
DO NOT allow end-user
to lie on an UNDER-INFLATED or OVER-INFLATED mattress overlay. Check the mattress overlay at least
once a day for proper adjustment.
NOTE: The mattress overlay is most effective when the Inflation Guidelines are followed and
there is air between all parts of the end-user and the base. Refer to the Inflation Guidelines
table to determine proper inflation level for areas of high risk. Under-inflation and over-inflation
reduce or eliminate the mattress overlay's therapeutic benefit, which may increase risk to the skin and
other soft tissue. The Inflation Guidelines table shows suggested distance from lowest bony prominence
to base of product for each section of the mattress overlay.
Inflation Guidelines:
Product
Distance From Base
Standard Overlay, Bariatric Overlay, O.R. Pad, and SOFFLEX 2
1/2 in. (1.5 cm) to 1 in. (2.5 cm)
PRODIGY
1 in. (2.5 cm) to 1.5 in. (4 cm)
NOTE: Insert Pads or Leveling Pads may be used as replacement of Standard Overlay sections ONLY.
DO NOT use them as replacement solutions where end-user is at risk of tissue breakdown. Replace
the pad(s) with Standard Overlay section(s) at first indication of tissue trauma. Ensure pad(s) and
sections(s) are properly snapped together before use.
LINENS
Too many linens between the mattress overlay and the end-user will compromise performance, prevent
the end-user from sinking in, and reduce pressure distribution. It is recommended that a maximum of:
• One flat or contoured sheet be arranged loosely over the mattress overlay.
• If needed, a draw sheet for moving and transferring can be used.
• If needed, one incontinent pad can be used. Water-resistant Reusable Overlay Covers and Overlay
Enclosure Covers are available from ROHO, Inc.
Special Positioning: If the end-user has special positioning needs while lying on his/her side or
back, a pillow can be placed beneath the mattress overlay. Consult with your clinician and / or
healthcare provider for proper pillow positioning.
PILLOW POSITIONING
CPR
For Standard Overlay, Bariatric Overlay, and O.R. Pad, CPR can be administered with or without a
cardiac board between the end-user and these mattress overlays. It is not necessary to deflate these
mattress overlays to achieve sufficient compression levels. ROHO, Inc. recommends the use of a
cardiac board for the SOFFLEX 2 and PRODIGY.