Safety Information

5

W

Warning

•

When using a pressure module, make sure the process pressure line is shut off and depressurized

before you connect it or disconnect it from the pressure module.

•

Use only 4 AA batteries, properly installed in the calibrator case, to power the calibrator.

•

Disconnect test leads before changing to another measure or source function.

•

When servicing the calibrator, use only specified replacement parts.

•

To avoid false readings, which could lead to possible electric shock or personal injury, replace the

battery as soon as the battery indicator (

M

) appears.

Caution

To avoid possible damage to calibrator or to equipment under test:

•

Disconnect the power and discharge all high-voltage capacitors before testing resistance or continuity.

•

Use the proper jacks, function, and range for your measurement or sourcing application.