Loop power supply, Rtd excitation (simulation), Temperature, rtd ranges, and accuracies (its-90) – Fluke 725 User Manual

Page 69

Advertising
background image

Specifications

59

Loop Power Supply

Voltage: 24 V

Maximum current: 22 mA

Short circuit protected

RTD Excitation (simulation)

Allowable Excitation by RTD type

Ni 120

0.15 to 3.0 mA

Pt 100-385

0.15 to 3.0 mA

Pt 100-392

0.15 to 3.0 mA

Pt 100-JIS

0.15 to 3.0 mA

Pt 200-385

0.15 to 3.0 mA

Pt 500-385

0.05 to 0.80 mA

Pt 1000-385

0.05 to 0.40 mA

Temperature, RTD Ranges, and Accuracies (ITS-90)

Accuracy

Type

Range

°

C

Measure 4-Wire

°

C

Measure 2- and 3-Wire*

°

C

Source

°

C

Ni120

-80 to 260

0.2

0.3

0.2

Pt100-385

- 200 to 800

0.33

0.5

0.33

Pt100-392

-200 to 630

0.3

0.5

0.3

Pt100-JIS

-200 to 630

0.3

0.5

0.3

Pt200-385

-200 to 250
250 to 630

0.2
0.8

0.3
1.6

0.2
0.8

Pt500-385

-200 to 500
500 to 630

0.3
0.4

0.6
0.9

0.3
0.4

Pt1000-385

-200 to 100
100 to 630

0.2
0.2

0.4
0.5

0.2
0.2

Resolution: 0.1

°

C, 0.1

°

F

Allowable excitation current (source): Ni120, Pt100-385, Pt100-392, Pt100-JIS, Pt200-385: 0.15 to 3.0 mA
Pt500-385: 0.05 to 0.80 mA; Pt1000-385: 0.05 to 0.40 mA
RTD Source: Addresses pulsed transmitters and PLCs with pulses as short as 5 ms.
* 2-wire: Does not include lead resistance.
3-wire: Assumes matched leads with a total resistance not exceeding 100

.