Diode test – Fluke 77 User Manual

Page 34

Advertising
background image

Maintenance

AC Voltage Test

3

3-9

Referring to Table 3-4, set the Ohms Calibrator or Decade Resistor to the
resistance value indicated in steps 1 through 6. Verify that the display reading
is within the limits shown.

3-10. Diode

Test

1.

Set the UUT to the Diode Test function.

Note

On Fluke 5100 series calibrators, activate the 50

Ω

divider

override. On Fluke 5500A calibrators, lock voltage to the 33V
range.

2.

Apply an input from the DC Voltage Calibrator of +.090V dc to the

z and COM input terminals of the UUT, and verify that the
beeper is on.

3.

Increase the DC Voltage Calibrator output to +.110V dc, and verify
that the beeper is off.