Intended use, English – Abbott Freestyle Freedom Lite User Manual

Page 5

Advertising
background image

1

Important Information About Your Blood Glucose Monitoring System

Intended Use

Use outside the body only (in vitro diagnostic use) .
The FreeStyle Freedom Lite® Blood Glucose Monitoring System is intended for use in the quantitative

measurement of glucose in capillary whole blood from the finger, upper arm and palm and venous

whole blood . It is intended for use by healthcare professionals and people with diabetes mellitus

at home as an aid in monitoring the effectiveness of a diabetes control program . It is not intended

for the diagnosis of or screening for diabetes mellitus, and it is not intended for use on neonates or

arterial blood .

IMPORTANT:

The device must not be used on more than one person including other family members due to the

risk of spreading infection . All parts of the device and its accessories are considered biohazardous

and can potentially transmit infectious diseases, even after performing the cleaning procedure .

1,2

Healthcare professionals handling this system must always wear gloves and should follow the

infection control policies and procedures approved by their facility .

Use only FreeStyle Lite® test strips and FreeStyle® control solution with this meter . Other products

can produce inaccurate results .

1

“FDA Public Health Notification: Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne

Pathogens: Initial Communication” (2010) http://www .fda .gov/MedicalDevices/Safety/AlertsandNotices/ucm224025 .htm .

2

“CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne Pathogens”

(2010) http://www .cdc .gov/injectionsafety/Fingerstick-DevicesBGM .html

Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com

English

Advertising