Nexen AWL 280 912662 User Manual

Page 12

Advertising
background image

12

FORM NO. L-20219-B-1212

CALIBRATION (continued)

RIGHT HAND SENSOR CALIBRATION.

Use an opaque sheet of material to completely block
light projection from the projector to the detector of the
right sensor (see Fig 8).

Connect the negative lead of a DMM to the "COM"
test pin on the main printed circuit board. Connect the
positive lead to the "OUT-R" test pin on the main printed
circuit board (See Fig. 9).

Adjust the "BIAS-R" trim pot until the DMM
indicates +4 ± 0.1 VDC (See Fig. 11).

Remove opaque material.

Adjust the "GAIN-R" trim pot until the DMM
indicates –4 ± 0.1 VDC (See Fig. 11).

If you can not achieve –4 ± 0.1 VDC, then increase the
intensity of the projector LEDs by adjusting the LED-R
trim pot

clockwise until –4 ± 0.1 VDC is achieved

and then recheck the sensor calibration.

NOTE: If your web application has required that you
increase the gap beyond the standard 1.95 [50 mm],
then you may need to increase the intensity of the
projector LEDs in order to achieve –4 ± 0.1 VDC.

Right HAND SENSOR ALARM CIRCUIT ADJUSTMENT.

Turn the "ALARM-R" trim pot fully counter clockwise.

The right LED on AS10-D front panel that is labeled
"Detector Dark" will turn "Off"

Turn the "ALARM-R" trim pot clockwise until the

LED comes "ON".

Turn the "ALARM-R" trim pot counter clockwise

slowly and stop when the LED goes out.

To test the circuit, insert four or five sheets of clear film
about 2 inches [50 mm] wide into the alarm sensor's
light path (See Fig. 10). The right LED on AS10-D front
panel that is labeled "Detector Dark" will turn "ON".

Right Hand Calibration

Refer to figures 8, 9, 10 and 11.