5 program security, 6 auto tune – ABUS Technologies A9696 Series Universal Controller User Manual

Page 19

Advertising
background image

ABUS TECHNOLOGIES INC.

19

A9696

7.4.8 FUNCTION KEY AND DIGITAL INPUT (I/O6) FUNCTIONS

Both the

F

key and the I/O6 digital input can be programmed to execute functions 7, 8, 9 and

10 shown in Table . The key function is configured in parameter “fFvn”. The digital input function is

configured in parameter IO6. The digital input can also be configured for function 6: Auto/Manual mode

change.

7.5 Program Security


Each menu cycle can be locked (protected) by pressing

and

simultaneously for 3

seconds. Press

and

for 3 seconds to unlock. A short blink of the display confirms the lock/unlock

change. This will alternately lock or unlock the

and

keys to avoid tampering.

For further protection, the unlock operation through the keypad may be disabled by changing

the position of an internal strap inside the controller:

When PROT is OFF, the user is allowed to lock and unlock the cycles using the keypad as

explained above. If PROT is ON, the cycles lock/unlock operation is disable.

7.6 Auto Tune

During auto tune the process is controlled in ON/OFF mode at the programmed Set-Point Value

(SV). Depending on the process characteristics large oscillations above and below SV may occur and

auto tuning may take several minutes to be concluded. The recommended procedure is as follows:

Disable the control output at the rvn prompt by selecting NO.

Select auto mode operation at the Avto prompt by selecting YES.

Disable the ramp and soak function (select NO) and program a new SV value other than the
present PV (close to the desired set point).

Enable auto tuning at the Atvn prompt by selecting YES.

Enable the control output at the rvn prompt by selecting YES.

During the auto tune procedure the soft-start function will not operate and large oscillations will

be induced around the setpoint. Make sure the process can accept these oscillations and fast control

output changes. If auto tuning results are not satisfactory refer to the table below for manual fine tuning

procedure.

Suggestions for manual tuning of PID parameters

PARAMETER

RESPONSE

SOLUTION

Proportional Band

Slow Response

Decrease

Large Oscillation

Increase

Integral Rate

Slow Response

Increase

Large Oscillation

Decrease

Derivative Time

Slow Response or

Instability

Decrease

Large Oscillation

Increase

Advertising